[口頭報告]A Raman Spectroscopic method for Measuring the Crystalline Silica Content in Coal Dust

A Raman Spectroscopic method for Measuring the Crystalline Silica Content in Coal Dust
編號:189 稿件編號:211 訪問權限:僅限參會人 更新:2024-05-19 11:24:00 瀏覽:2137次 口頭報告

報告開始:2024年05月30日 19:30 (Asia/Shanghai)

報告時間:10min

所在會議:[S2] Safety Engineering and Occupational Health ? [S2-4A] Evening of May 30th-4A

暫無文件

摘要
To explore the use of Raman spectroscopy in analyzing crystalline silica content within coal dust, the Raman spectroscopy techniques were investigated for quantitative analysis of silica in coal dust samples from China based on unary linear regression (ULR) and partial least squares regression (PLSR) method. In the experimental setup, the formulated coal dust samples with known crystalline silica content were meticulously prepared and subjected to an ashing procedure using a muffle furnace. The cooled mixtures were placed in water for magnetic stirring before being dispensed onto an aluminum substrate via a pipette gun, ultimately forming dry sample deposits. According to the Raman spectrum data from these dry sample deposits, a calibration curve was generated and served as the foundation for predicting crystalline silica content in real coal dust samples. In the quantitative analysis of the crystalline silica load in the formulated coal dust samples deposited within a diameter of about 1 cm, the calibration curve R2 value of the ULR model is 0.90, with a limit of detection (LOD) is 0.96% (i.e., 19.17 μg). Meanwhile, the R2 value of the PLSR model can reach 0.988, and concurrently, the LOD can be reduced to 0.18% (3.65 μg). Moreover, the improved PLSR model, which established with coal dust samples collected from a broader range of geological conditions, demonstrates high accuracy and good agreement between the analysis results of the prediction set and the measurements obtained from the standard infrared (IR) spectrometric method. This study indicates the potential of Raman spectroscopy for measuring crystalline silica in coal dust.
關鍵字
coal dust,crystalline silica,quantitative analysis,Raman spectroscopy
報告人
Wenting Feng
China University of Mining and Technology